Basic Info.
Measurement Project
All Types of Gasoline, Diesel
Octane Number Measurement in Line
ASTM D2699-86
Diesel Cetane Number Measurement in Line
ASTM D4737-03
Measuring Range
Octane Number :40-120
Each Sample Measurement Time
<8 Seconds
Instrument Current Consumption
30mA
Battery Standby Time
100 Hours
Operating Range
Temperature -10 ºC to +40 ° C
Relative Humidity
30% to 81% R.H
Atmospheric Pressure
64 ~ 106kpa
Transport Package
Wooden Case
Specification
236mmx130mmx43mm
Product Description
Instrument introduction
OC130 is a portable, digital display octane cetane number tester, it can be quick and easy, accurate and not damage the oil sample to detect octane number of gasoline and cetane number of diesel, applies to the laboratory, can also be used for site inspection. It is the domestic first instrument which can not only detect the gasoline octane number, can also be applied to diesel cetane number. It uses dielectric constant method to detect octane number and cetane number detection technology, fully in line with international standards (ASTM D2699-86, ASTM D2700-86), commonly used in motor gasoline and diesel quality analysis, and correspond to study method RON, motor octane number MON and cetane number CN.
Main features
Measure gasoline octane number and diesel cetane number.
With the most authoritative gasoline octane and diesel cetane number database
The repeatable error range is 0.2 octane units.
Measuring gasoline containing additives.
Display RON, MON and antiknock index (AKI).
Measuring diesel cetane number CN, diesel type.
Powerful processing chip can quickly and accurately process the data.
The instrument is equipped with six kinds of oil test data model to adapt to a variety of complex conditions under oil measurement.
Easy operation, small size, easy to carry, cabinet vibration-proof, anti-solvent, sealing, low-cost.
Can be connected to the computer by RS232 data transmission lines, you can directly operate and print by the computer.
Traditional octane number test inspection machine and cetane number testing machine requires professional and technical personnel to operate, and the process is tedious, and need to consume a large number of reference fuel, the equipment maintenance cost is expensive. There are great difficulties in determination of octane number and cetane number for ordinary users. The efficiency of Lab130 fuel measurement method is higher than that of traditional method.
Measurement principle
Octane number detector principle lies in the insulating permeability and charge characteristics of the electromagnetic induction of the octane number of petrol and cetane number of diesel to measure. By measuring the dielectric properties of the oil, compared to the data model of known existence in the memory, thereby determining the results. The instrument is very sensitive, and can measure the tiny dielectric parameters change to get ctane number, cetane number, and other petroleum products parameters.
Technical parameters
Measurement project: all types of gasoline, diesel
Octane number measurement in line: ASTM D2699-86, ASTM D2700-86.
Diesel cetane number measurement in line: ASTM D4737-03, ASTM D613, EN ISO 5165.
Measuring range: octane number :40-120 (accuracy ± 0.5), cetane number :20-100 (accuracy ± 1)
Each sample measurement time: <8 seconds
Instrument current consumption: 30mA,
Battery standby time: 100 hours
Operating Range: temperature -10 ºC to +40 ° C, relative humidity 30% to 81% R.H., atmospheric pressure 64 ~ 106kPa
Dimensions
Sensor: 60mm × 100mm,
Host machine: 236mmx130mmx43mm
OCTANE & CETANE ANALYZER |
Measurement project | all types of gasoline, diesel |
Octane number measurement in line | ASTM D2699-86, ASTM D2700-86. |
Diesel cetane number measurement in line | ASTM D4737-03, ASTM D613, EN ISO 5165. |
Measuring range: octane number | 40-120 (accuracy ± 0.5) |
Each sample measurement time | <8 seconds |
Instrument current consumption: | 30mA |
Address:
The 9th Floor, District A, Liancheng Waiking Street, Jianshe Road, Yuetang, Xiangtan, Hunan, China
Business Type:
Manufacturer/Factory, Trading Company
Business Range:
Health & Medicine, Industrial Equipment & Components, Instruments & Meters
Management System Certification:
ISO 9001, ISO 9000, ISO 14001, ISO 14000, ISO 14064, GMP
Company Introduction:
As professional laboratory equipment manufacturer, we are in line with experimental equipment for many years, such as Vapor Permeability Tester, Gas permeation rate tester, drying oven, testing machine, Purification equipments, Geology testing, ICP&DRS, etc. <br><br>We have 20 years eexperience in this industry. We have sophisticated technology to serve for enterprise, scientific research institution, universities and colleges, etc. We cooperate with many research institution and university, etc. <br><br>Our product is widely used in Europe, North America, Asia, Africa, Latin America, etc. We have branches in hunan, and HK. We are paying attention to service and improving our product these years. Customer is everything, which is our service aim; All of our efforts are focused on providing the suitable testing instruments with reliable results. We believe that high quality and good service will pay important role in the future. <br><br>We stick to the principle of "quality first, service first, continuous improvement and innovation to meet the customers" for the management and "zero defect, zero complaints" as the quality objective. To perfect our service, we provide the products with the good quality at the reasonable price. <br><br>We have persisted on the business idea of "Integrity Basis and Quality First" for the stable and constant development. We are making effort to develop together with customers from both domestic and abroad on the basis of reciprocity. Client, quality and service are our supreme target!