Vapor Permeability Tester, Permeability Tester, Differential Pressure Method Gas Permeameter manufacturer / supplier in China, offering Air Permeability Tester for Rubber, Portable Gas Detector for Carbonic Oxide and Carbon Dioxide, Freezing Point of Aviation Fuels and Cloud & Crystallization Point and so on.

(/ )

Supplier Homepage Product Packing test Gas permeation rate tester Air Permeability Tester for Rubber

Air Permeability Tester for Rubber

FOB Price: Get Latest Price
Min. Order: 1 Piece
Port: Shenzhen, China
Production Capacity: 1000 Sets Per Month
Payment Terms: L/C, T/T, Western Union, Paypal, Money Gram

Request a custom order and have something just for you!

Send Customized Request
Basic Info
  • Model NO.: LA-APT-33D
  • Test Precision: 0.01 Cc/M2•day•0.1MPa
  • Primary Sensor Quantity: 3
  • Vacuum Error: 0.1PA
  • Vacuum Efficiency: Less Than 27PA in 5 Minutes
  • Humidity Precision: ±1%Rh
  • Transport Package: Cartons
  • Origin: China
  • Measurement Range(Film): 0.05-100,000
  • Test Sample Quantity: 3
  • Test Pattern: Independent Sensor
  • Vacuum Degree: <10PA
  • Temperature Control Accuracy: ±0.1ºC
  • Trademark: labmen, OEM
  • Specification: 880× 610× 430mm
  • HS Code: 9031809090
Product Description
Product function:
It is applied to various gas permeation rate detection of O2, CO2, N2 and air, etc. Also in plastic film, composite membrane, High-barrier materials, sheet, metal foil, rubber, etc. 

Operating principle:
Using the principle design of differential pressure method, put treated sample in the middle of upper and bottom chamber in advance. Pressure tightly and seal,pick up the suction to the whole system, after reaching the required vacuum degree, close bottom chamber, fill with certain test gas to high pressure chamber, and adjust upper chamber pressure, keep on two sides of the sample forming constant air pressure gradient difference, the gas is under the action of differential pressure gradient, permeating from high-pressure side (upper chamber) to vacuum side (bottom chamber).Accurately measure vacuum chamber (bottom chamber) intensity of pressure variation, calculate various barrier property parameters of test sample. Fully conform to the standard ASTM D1434.
 
Application characteristic
High speed vacuum ability, to reduce the vacuum time,and accelerate the degassing, reduce test error;
Short test time, high efficiency;
Could extend to test bottle, bag;
Less gas consumption, gas bottle can be used continuously for one month;
Can measure any gas, including toxic gases, no gas leak, and the gas consumption up to six months a bottle;
Automatically adjust the test pressure; to adapt to the a variety of pressure requirements, can be as low as 2kPa;
Power failure protection;
A whole-process monitor, automatic recording, adjustable sampling rate,
Reappear in the whole test process again. Strong graph curve function, convenient for comparison test and analysis;
Automatic testing: Set the test parameters, test automatically proceeds after starting, automatically judge the test is over and save the test result.
There is general data base system in host software, data could be exported, could store the data, inquire, print, etc. Through the analysis of historical data,it's easy to grasp the history dynamic of the products.
 
Product feature
Pure semiconductor temperature control, automatically control heating and cooling,Temperature range is wider and more accurate,parallel work, higher reliability.
Apply base line technology; remove the system error, test result is more accurate.
Be of accurate pressure control ability, maintaining constant pressure difference.
International top vacuum sensors, high-quality components, precision piping system to ensure the stability and reliability of the instrument operation,
Embedded system control, fully controlled by computer.
 
 Meeting Standards
Compatible with the standard of various tests:ASTM D1434,ISO 2556,ISO 15105-1,JIS K7126-A,YBB 00082003

Technical index

Technical parameter

Model

LA-APT-33D

Measurement range(film)
(cc/m2·day·0.1MPa)

0.05-100,000

test precision

0.01 cc/m2•day•0.1MPa

Test sample quantity

3

Primary sensor quantity

3

test pattern

Independent sensor

vacuum error

0.1Pa

vacuum degree

<10Pa

Vacuum efficiency

Less than 27Pa in 5 minutes
Less than 20Pa in 10 minutes

testing efficiency

Test time < 4 hours, when transmittance >5cc/m2.day.0.1MPa

Temperature range

5ºC-55ºC (90ºC is optional.)

temperature control accuracy

±0.1ºC

humidity range

5~90 %RH(double-flowhumidity method)

Humidity precision

±1%RH

test pressure

0.001MPa~0.2Mpa

air source pressure

0.1MPa~0.8 MPa

test gas

arbitrary gas (gas discharge port customers-owned)

carrier gas port

         ¢8

vacuum tube diameter

¢15

Display control mode

fully controlled by computer

Correction mode

Baseline correction, Standard film correction,vacuum transducer external correction

sample thickness

≤3mm

Test area

50cm2, circle (other dimension could be customized.)

Main machine dimension
(L*B*H)

880×610×430mm

Weight(KG)

40kg

60kg

65Kg

Power supply

220V AC 50HZ

 Air Permeability Tester for Rubber

 Air Permeability Tester for Rubber
Air Permeability Tester for RubberAir Permeability Tester for RubberAir Permeability Tester for Rubber
Air Permeability Tester for Rubber

 
Send your message to this supplier
Avatar
Mr. David
Marketing Manager
*From:
To: Mr. David
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now